Get Your FREE Helium Hotspot

Wednesday, March 18, 2009

Membrane Nitrogen Generators

Parker Balston® Membrane Nitrogen Generators are designed to supply single or multiple LC/MS instruments with dry nitrogen at purities of 99% to 99.5%. The generator can also be used for solvent evaporation as well as supplying dry nitrogen to analytical instruments.

Low Flow Membrane N2 Generators
NitroFlow (includes compressor): 30 lpm nominal
N2-04: 4 slpm
N2-14, N2-14ANA: 14 slpm

Mid Flow Membrane N2 Generators
N22, N2-22ANA: 22 slpm
N2-35, N2-35ANA: 35 slpm

High Flow Membrane N2 Generators
N2-45, N2-45A: 45 lpm
N2-80, N2-80ANA: 80 lpm
N2-135, N2-135ANA: 135
NitroVap 1LV: 80 slpm
NitroVap 2LV: 160 slpm

FT-IR Purge Gas Generators

Parker Balston FT-IR Purge Gas Generators are specifically designed for use with FT-IR Spectrometers to provide a purified purge gas and air bearing gas from compressed air.

The generators supply carbon dioxide-free air at less than -100°F (-73°C) dew point with no suspended impurities larger than 0.01 micron. The units are designed to operate continuously 24 hours a day, 7 days a week. The Parker

Balston FT-IR Purge Gas Generator completely eliminates the inconvenience and the high costs of nitrogen cylinders and dewars, and significantly reduces the costs of operating all FT-IR instrumentation. Each system offers cleaner background spectra in a shorter period of time and more accurate analysis by improving the signal-to-noise ratio. The typical payback period is less than one year. The generators are also ideally suited for use with CO2 Analyzers in addition to supplying other laboratory instruments. Each generator is quiet, reliable, and easy to install - simply attach the inlet air and outlet purge lines, plug the power cord into a wall outlet, and enjoy trouble-free operation. The Model 74-5041NA Self Contained FT-IR Purge Gas Generator includes a state-of-the-art compressor.

Specifications:

Flow Rates @ Inlet Pressure of 125 psig and (60 psig):
Model 75-45NA: 36 scfh (18 scfh)
Model 75-52NA: 72 scfh (36 scfh)
Model 75-62NA: 216 scfh (120 scfh)

CO2 Concentration (all models): < 1ppm

Min/Max Inlet Air Pressure (all models): 60 psig / 125 psig

GC Gas Station

The Model GCGS-7890NA GC Gas Station can provide both hydrogen and zero grade air to FID detectors on GCs. This system is designed to provide carrier, fuel gas and support air to 1-3 FIDs and 1-3 capillary columns.

The hydrogen cell produces up to 500 cc/min of 99.99999+% pure hydrogen gas. The zero air compartment produces up to 3500 cc/min or zero grade air.

The GC Gas Station is a complete system with state-of-the-art, highly reliable components engineered for easy installation, operation, and long term performance.

Specifications:

Hydrogen Purity: 99.99999+%
Zero Air Purity: <0.05 ppm (total hydrocarbons as methane)
Maximum Hydrogen Flow Rate: 55 cc/min
Maximum Zero Air Flow Rate: 3500 cc/min
Electrical Requirements: 120 VAC, 60 Hz, 5.3 Amps
Hydrogen Outlet Pressure: 100 psig
Zero Air Outlet Pressure: 40-125 psig
Dimensions: 11"w x 27"d x 17"h (28cm x 69cm x 43cm)
Inlet Port: 1/4" NPT (female tube) compressed air supply
Outlet Ports: 1/8" Compression, Stainless
Shipping Weight: 60 lbs / 27 kg